Key Features of the Automatic Kjeldahl Nitrogen Analyzer
- Measuring range 0.1 to 240 mg nitrogen
- Recovery of 99.5 percent or better, repeatability ±0.5%
- Automatic addition of alkali, dilution water and receiving solution
- Colorimetric titration and result calculation on advanced models
- Safety door, cooling water monitoring and automatic fault detection
- Data storage up to 140,000 results on YSKN-987 models
Automatic Kjeldahl Nitrogen Analyzer Technical Specifications
| Model | Yskn-984 | Yskn-110F | Yskn-986 |
|---|---|---|---|
| Water Consumption | 1.5L/Min | ||
| Ambient Temperature | 5~40℃ | ||
| Ambient Humidity | 80% | ||
| Measure Range | 0.1mg~200mg | ||
| Analysis Time | 5~10min | ||
| Recovery Ratio | ≥99.5% | ||
| Sample Amount | Solid ≤6g; Liquid ≤16ml | Solid ≤ 5g; Liquid ≤20ml | |
| Burette Accuracy | / | 1.0μL/ step | 2μL/ step |
| Reproducibility | / | ±0.5% | ±0.5% |
| Consumption | 2000W | ||
| Power Supply | AC110/220 V ±10%,60/50Hz | ||
| Packing size(W*D*H) | 655*528*880mm | Main body:585*525*900mm Accessories:605*600*355mm | Main body:585*525*905mm Accessories:600*600*360mm |
| Gross weight | 45Kg | Main body:52.7kg, Accessories:12.5kg | Main body:50.3Kg, Accessories:13.4kg |
| Functions List | Yskn-984 | Yskn-110F | Yskn-986 |
|---|---|---|---|
| Variable Steam Generator | ● | ● | ● |
| Cooling Water Pressure Monitor | ● | ● | ● |
| Cooling Water Temperature Monitor | ● | ● | |
| Automatic Alkali Solution Adding | ● | ● | ● |
| Automatic Dilute Water | ● | ● | ● |
| Auto Solution Adding | ● | ● | ● |
| Tube Emptying | ● | ● | ● |
| Removable Drain Tray | ● | ● | ● |
| Safety Door | ● | ● | ● |
| Tube on Position Sensor | ● | ● | |
| Tube Replacement Sensor | ● | ● | |
| Colorimetric Titration Device | ● | ● | |
| Result Calculation | ● | ● | |
| Result and Data Storing | ● | ● | |
| Built-in Printer | ● | ||
| Automatic fault detection function | ● | ● | ● |
| Titration While Distillation | ● |
| Model | Yskn-987 | Yskn-987A |
|---|---|---|
| Working Process | Adding quantitative sample dilution, alkali liquor and absorption liquid; Automatically finishing distillation, titration, print and liquid-discharging. | |
| Measurement Range | 0.1~240mgN | |
| Sample Amount | Solid<6g; liquid<16ml | |
| Measurement Time | 4~6 minutes | |
| Titration Precision | 0.75μL/step | |
| Requirement of Cooling Water | <18℃ | |
| Cooling Water Consumption | 1.5L/min (Distillation process) | |
| Titration Solution Concentration | Direct titration samples with 4 titration concentrations of 0.1mol,0.5 mol and 1 mol | |
| Titration System | High-precision plunger type titration system | High-precision peristaltic titration system |
| Repeatability | ±0.5% | |
| Recovery Ratio | ≥99.5% | |
| Data Storage | 140,000 groups | |
| Consumption | 2000W | |
| Power Supply | AC220V, 50/60Hz | |
| Packing Size(W*D*H) | 510*460*710mm | |
| Gross Weight | 36.4Kg | 39.8Kg |
Specifications can vary by model and configuration. The final specification is confirmed in your quotation.
Applications
- Food and feed protein testing
- Dairy and meat analysis
- Fertilizer and soil nitrogen
- Water and wastewater (TKN)
- Pharmaceutical and agricultural research
Standards and Compliance
The Kjeldahl method is the reference method for protein in many food and feed standards, including AOAC, ISO and BIS methods. Samples are first digested in a digestion block, then distilled and titrated in this analyzer.
How to Choose the Right Automatic Kjeldahl Nitrogen Analyzer
Choose YSKN-984 for distillation only, YSKN-110F or YSKN-986 for distillation with built in titration, and YSKN-987 or 987A for fully automatic operation with large data storage.
Frequently Asked Questions
What does a Kjeldahl nitrogen analyzer measure?
It measures total nitrogen in a digested sample. The ammonia released during distillation is captured and titrated, and the nitrogen value is multiplied by a protein factor, such as 6.25 for many foods, to calculate crude protein content.
How long does one Kjeldahl analysis take?
Distillation and titration take about 4 to 10 minutes per sample on Laboteck analyzers, depending on the model. Digestion is a separate step that usually takes 1 to 2 hours in a digestion block before the sample goes to the analyzer.
Which model should I choose?
For basic distillation choose YSKN-984. For automatic titration and result calculation choose YSKN-110F or YSKN-986. For full automation, large data storage and a choice of plunger or peristaltic titration, choose YSKN-987 or YSKN-987A.
What is the price of Automatic Kjeldahl Nitrogen Analyzer in India?
The Automatic Kjeldahl Nitrogen Analyzer price depends on the model, size, configuration and accessories you choose, so Laboteck quotes each requirement individually instead of publishing a fixed price list. Send your requirement through the quote form or WhatsApp and you will receive a quotation in INR with GST, delivery time and warranty terms.
Do you export Kjeldahl analyzers?
Yes. Laboteck supplies Kjeldahl nitrogen analyzers to laboratories across India and exports to buyers worldwide. Share the model or size you need, quantity, power supply and destination country, and our team will reply with a quote in INR or USD, delivery time and shipping options.